IntelePACS Software (Intelerad) – image calculation error (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
IntelePACS (Image Fusion Module) - InteleViewer
Brand
INTELERAD MEDICAL SYSTEMS INCORPORATED 800, Boul. de Maisonneuve Est 12th floor Montreal Canada
Lot Codes / Batch Numbers
UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later
Products Sold
UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later
INTELERAD MEDICAL SYSTEMS INCORPORATED 800, Boul. de Maisonneuve Est 12th floor Montreal Canada is recalling IntelePACS (Image Fusion Module) - InteleViewer due to A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Stan. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Recommended Action
Per FDA guidance
On 2/5/2025, correction notices were emailed to customers who were asked to do the following: The following work around can be completed for any impacted SUV calculation. The correct SUV values may be obtained in the SUV calculator, as follows: 1. Adjust the administration time and series time in IntelePACS consistent with the actual time of injection and the time of series acquisition. 2. Verify that the calculated elapsed time is correct. 3. Recalculate the SUV in the calculator, this is the correct SUV value. - Customers are asked to review scans and notify the recalling firm if scans were performed on Daylight Savings Time dates since 2022. - A software upgrade to eliminate this bug is expected to be released by the end of March 2025. Once available, you will be notified via email. You will then be able to schedule your product upgrade. - Complete and return the acknowledgement form via email to regulatory-affairs@intelerad.com For more information or to ask any questions about this notification, please contact the firm's regulatory affairs department at regulatory-affairs@intelerad.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.