Intuitive Surgical, Inc. Intuitive brand Da Vinci S Surgical System, an Endoscopic Instrument Control System, Model SS2000 Patient Side Cart, Model PS2000, Part Number: 380267-01, IS2000, Top Level, 4-Arm; Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Intuitive brand Da Vinci S Surgical System, an Endoscopic Instrument Control System, Model SS2000 Patient Side Cart, Model PS2000, Part Number: 380267-01, IS2000, Top Level, 4-Arm; Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA
Brand
Intuitive Surgical, Inc.
Lot Codes / Batch Numbers
System Numbers: SG011, SG012, SG013, SG014, SG017, SG018, SG019, SG020, SG058, and SG094.
Products Sold
System Numbers: SG011, SG012, SG013, SG014, SG017, SG018, SG019, SG020, SG058, and SG094.
Intuitive Surgical, Inc. is recalling Intuitive brand Da Vinci S Surgical System, an Endoscopic Instrument Control System, Model SS2000 Pa due to Da Vinci S Surgical System's spinal pin could limit mechanical motion of the arm and render the system unavailable for surgical use, and may result in. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Da Vinci S Surgical System's spinal pin could limit mechanical motion of the arm and render the system unavailable for surgical use, and may result in system failure and the need to convert to an alternative surgical technique.
Recommended Action
Per FDA guidance
The firm did not release notification to its consignees. The reworks were arranged as part of routine service visits initiated on 4/12/06.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026