Invacare Corp Tarsys 1st Generation Seating System - an optional accessory for the Powered Wheelchairs. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Tarsys 1st Generation Seating System - an optional accessory for the Powered Wheelchairs.
Brand
Invacare Corp
Lot Codes / Batch Numbers
All Model Numbers (excluding the 2nd generation models), Serial Numbers: 93I through 02E
Products Sold
All Model Numbers (excluding the 2nd generation models); Serial Numbers: 93I through 02E
Invacare Corp is recalling Tarsys 1st Generation Seating System - an optional accessory for the Powered Wheelchairs. due to A short circuit within the charger harness may cause heat damage to the units with potential for fire.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A short circuit within the charger harness may cause heat damage to the units with potential for fire.
Recommended Action
Per FDA guidance
On 3/10/2003, the firm mailed recall letters to all affected dealer accounts. The dealers were asked to contact all customers who have purchased powered wheelchairs with the optional Tarsys Seating System.