Invacare Corporation Invacare Storm TDX SR Power Wheelchair with Stability Lock Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-HD, TDXSR-MCG, TDXSR-MCG-HD, TDXSRV-HD Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Invacare Storm TDX SR Power Wheelchair with Stability Lock Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-HD, TDXSR-MCG, TDXSR-MCG-HD, TDXSRV-HD
Brand
Invacare Corporation
Lot Codes / Batch Numbers
Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-HD, TDXSR-MCG, TDXSR-MCG-HD, TDXSRV-HD. Serial Numbers: 08BE000082, 08BE000088, 08BE000089, 08BE000090, 08BE000323, 08BE001102, 08BE001105, 08BE002283, 08BE002803, 08BE003025, 08BE003250, 08BE003251, 08BE003472, 08BE003473, 08BE003632, 08BE004402, 08BE004414, 08BE004415, 08BE004835, 08BE005297, 08DE002040, 08EE000277, 08EE001539, 08EE001540, 08EE001775, 08FE003254, 08FE003257, 08FE003668, 08FE004681, 08GE003570, 08GE003936, 08HE001169, 08HE005013, 08IE001081, 08IE001316 & 08IE002729.
Products Sold
Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-HD, TDXSR-MCG, TDXSR-MCG-HD, TDXSRV-HD. Serial Numbers: 08BE000082; 08BE000088; 08BE000089; 08BE000090; 08BE000323; 08BE001102; 08BE001105; 08BE002283; 08BE002803; 08BE003025; 08BE003250; 08BE003251; 08BE003472; 08BE003473; 08BE003632; 08BE004402; 08BE004414; 08BE004415; 08BE004835; 08BE005297; 08DE002040; 08EE000277; 08EE001539; 08EE001540; 08EE001775; 08FE003254; 08FE003257; 08FE003668; 08FE004681; 08GE003570; 08GE003936; 08HE001169; 08HE005013; 08IE001081; 08IE001316 & 08IE002729.
Invacare Corporation is recalling Invacare Storm TDX SR Power Wheelchair with Stability Lock Model numbers: TDXSR, TDXSR-CG, TDXSR-CG- due to For some Invacare chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage prope. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
For some Invacare chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage properly, the wheelchair may do one or more of the following: (1) veer to one side; (2) rock forward onto its front riggings; (3) drive in an unintended circular motion; and/or (4) fall forward or to one side. In these cases there is a risk to injury for
Recommended Action
Per FDA guidance
On 9/15/2009 the firm sent recall notification letter to their customers instructing them to immediately contact the dealer to have the stability lock inspected and/or adjusted. If the dealer is unavailable or the repair will take longer than 60 days, they are to contact the firm and arrange for an inspection and/or adjustment. Invacare requested contact names to whom it can send additional information. Invacare plans to mail, via FEDEX, information regarding the recall to each consignee. Initial consignees received a response card and instructions to forward cards to any customer to whom they have distributed the product. The products are to be inspected at the cylinder label and if the label does not have a long alpha-numeric string demonstrated in the label, the cylinders on the chair have to be replaced. Questions are directed to Recalls@Invacare.com or to 800-333-6900.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AZ, CA, CO, CT, FL, GA, IL, IN, IA, KS, ME, MA, MI, MN, MS, MO, NM, NY, NC, ND, OH, SC, TX, DC