Inverness Medical West Nile Virus IgM Capture ELISA The Panbio West Nile Virus IgM Capture ELISA is for the qualitative presumptive detection of IgM antibodies to West Nile virus in serum as an aid in the clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis/meningitis. Positive results must be confirmed by plaque reduction neutralization test (PRNT), or by using the current Centers for Disease Control and Prevention (CDC) guidelines for diagno Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
West Nile Virus IgM Capture ELISA The Panbio West Nile Virus IgM Capture ELISA is for the qualitative presumptive detection of IgM antibodies to West Nile virus in serum as an aid in the clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis/meningitis. Positive results must be confirmed by plaque reduction neutralization test (PRNT), or by using the current Centers for Disease Control and Prevention (CDC) guidelines for diagno
Brand
Inverness Medical
Lot Codes / Batch Numbers
Lot 07262 (Canada), and Lot 08118 (Four customers in USA)
Products Sold
Lot 07262 (Canada), and Lot 08118 (Four customers in USA)
Inverness Medical is recalling West Nile Virus IgM Capture ELISA The Panbio West Nile Virus IgM Capture ELISA is for the qualitati due to It has come to our attention that users of these product lots may experience an elevation in the rate of false positivity when compared with previous . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
It has come to our attention that users of these product lots may experience an elevation in the rate of false positivity when compared with previous product lots. Internal quality assurance testing has shown that the products are performing within the release specifications, although there has been a slight shift in the cut-off which may lead to an increase in the number of positive results obta
Recommended Action
Per FDA guidance
Inverness Medical sent an Urgent In Vitro Diagnostic Recall letter dated October 6, 2008, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use, destroy any unused or partially used test kits of the affected lot(s), and complete the table and certification of compliance. Customers were asked to fax back the completed form to inverness Medical at 1-609-527-8054. Replacement stock would be forwarded to them at no cost when the new lot becomes available. For questions regarding this recall call 609-627-8038.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
MN, ND, PA, TX