Iris Sample Processing StatSpin Express 3 Centrifuge with RTX8 Rotor Model: M502 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
StatSpin Express 3 Centrifuge with RTX8 Rotor Model: M502
Brand
Iris Sample Processing
Lot Codes / Batch Numbers
Rotors with serial numbers: B1562 to B1738
Products Sold
Rotors with serial numbers: B1562 to B1738
Iris Sample Processing is recalling StatSpin Express 3 Centrifuge with RTX8 Rotor Model: M502 due to Rotor may crack and separate causing device to fail. A failed rotor not successfully contained may seriously injure the laboratory worker.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Rotor may crack and separate causing device to fail. A failed rotor not successfully contained may seriously injure the laboratory worker.
Recommended Action
Per FDA guidance
Iris telephoned customers on 9/29/06 and advised them of the recall of the rotor. New rotors were mailed Overnight Confirmed Delivery with instructions for replacement and return of the the old rotor.