Isopure Corp Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3
Brand
Isopure Corp
Lot Codes / Batch Numbers
ACDS UDI: (01)00852456007092, ACDS Serial Number AC-837874 with Pump SN 178745C18, ACDS Serial Number AC-834638 with Pump SN 180212D18, ACDS Serial Number AC-837654 with Pump SN 178324B18, ACDS Serial Number AC-837864 with Pump SN 178743C18, ACDS Serial Number AC-837664 with Pump SN 178322B18, ACDS Serial Number AC-838050 with Pump Serial Number 178752C18, ACDS Serial Number AC-838030 with Pump SN 178746C18, ACDS Serial Number AC-838020 with Pump SN 178748C18, ACDS Serial Number AC-838173 with Pump SN 178744C18, ACDS Serial Number AC-838143 with Pump SN 178325B18, ACDS Serial Number AC-838260 with Pump SN 180211D18, ACDS Serial Number AC-838740 with Pump SN 180209D18, ACDS Serial Number AC-838743 with Pump SN 180207D18, ACDS Serial Number AC-838749 with Pump SN 180215D18, ACDS Serial Number AC-838746 with Pump SN 180206D18
Products Sold
ACDS UDI: (01)00852456007092; ACDS Serial Number AC-837874 with Pump SN 178745C18; ACDS Serial Number AC-834638 with Pump SN 180212D18; ACDS Serial Number AC-837654 with Pump SN 178324B18; ACDS Serial Number AC-837864 with Pump SN 178743C18; ACDS Serial Number AC-837664 with Pump SN 178322B18; ACDS Serial Number AC-838050 with Pump Serial Number 178752C18; ACDS Serial Number AC-838030 with Pump SN 178746C18; ACDS Serial Number AC-838020 with Pump SN 178748C18; ACDS Serial Number AC-838173 with Pump SN 178744C18; ACDS Serial Number AC-838143 with Pump SN 178325B18; ACDS Serial Number AC-838260 with Pump SN 180211D18; ACDS Serial Number AC-838740 with Pump SN 180209D18; ACDS Serial Number AC-838743 with Pump SN 180207D18; ACDS Serial Number AC-838749 with Pump SN 180215D18; ACDS Serial Number AC-838746 with Pump SN 180206D18
Isopure Corp is recalling Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifug due to There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue coul. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Recommended Action
Per FDA guidance
The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, FL, GA, MD, MA, MN, NM, NY, NC, OH, OK, TX, VA