Isopure Corp Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)
Brand
Isopure Corp
Lot Codes / Batch Numbers
Pump Serial Numbers: 178323B18, 178751C18, 178749C18, 180214D18, 180213D18, 180208D18, 178747C18, 178750C18
Products Sold
Pump Serial Numbers: 178323B18, 178751C18, 178749C18, 180214D18, 180213D18, 180208D18, 178747C18, 178750C18
Isopure Corp is recalling Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part) due to There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue coul. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Recommended Action
Per FDA guidance
The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.