Kamiya Biomedical Company, LLC K-ASSAY Alpha-1 Microglobulin Reagent Kit. Kit contains: 1 vial Reagent 1, Buffer 50 mL, and 1 vial reagent 2, Antibody Reagent 50 mL. The product is packaged in a kit box along with with the package insert. The package insert is labeled in part: "K-ASSAY Alpha-1 Microglobulin...Cat. No. KAI-056...For the Quantitative Determination of human Alpha-1 Microglobulin in urine, serum, and plasma...KAMIYA BIOMEDICAL COMPANY...12779 Gateway Drive...Seattle, WA 98168 USA...". Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
K-ASSAY Alpha-1 Microglobulin Reagent Kit. Kit contains: 1 vial Reagent 1, Buffer 50 mL, and 1 vial reagent 2, Antibody Reagent 50 mL. The product is packaged in a kit box along with with the package insert. The package insert is labeled in part: "K-ASSAY Alpha-1 Microglobulin...Cat. No. KAI-056...For the Quantitative Determination of human Alpha-1 Microglobulin in urine, serum, and plasma...KAMIYA BIOMEDICAL COMPANY...12779 Gateway Drive...Seattle, WA 98168 USA...".
Brand
Kamiya Biomedical Company, LLC
Lot Codes / Batch Numbers
Lot number 418041, catalog number KAI-056.
Products Sold
Lot number 418041, catalog number KAI-056.
Kamiya Biomedical Company, LLC is recalling K-ASSAY Alpha-1 Microglobulin Reagent Kit. Kit contains: 1 vial Reagent 1, Buffer 50 mL, and 1 vial due to Standardization of the reagent performance data did not match the standardization used to assign the calibrator values.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Standardization of the reagent performance data did not match the standardization used to assign the calibrator values.
Recommended Action
Per FDA guidance
Kamiya Biomedical Company, LLC sent an urgent medical Device Recall letter dated September 17, 2008, and the corrected package insert via e-mail. . The letter identified the product, the problem, and the action the customer should take. Customers were instructed to replace their previous package inserts with the corrected version and asked to sign and FAX the letter back once that had been completed. As one customer is also a distributor, the recalling firm told the this distributor's representative to advise their single customer of this correction and replace the package insert, and return the signed letter back to the firm. On September 22, 2008, both customers were contacted by phone and reminded to fax the signed letters once completed. For questions regarding this recall call 1-800-222-0342.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
IN, MN, PA