Kamiya Biomedical Company, LLC "K-ASSAY CRP (3) Calibrator E". This product is an IVD calibrator for use on automated clinical chemistry analyzers by laboratories. It is used for the calibration of the K-ASSAY CRP (3) immunoturbidimetric assay for quantifying C-Reactive Protein (CRP) in serum and plasma (sodium EDTA or lithium heparin) samples. It is for in vitro diagnostic use. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
"K-ASSAY CRP (3) Calibrator E". This product is an IVD calibrator for use on automated clinical chemistry analyzers by laboratories. It is used for the calibration of the K-ASSAY CRP (3) immunoturbidimetric assay for quantifying C-Reactive Protein (CRP) in serum and plasma (sodium EDTA or lithium heparin) samples. It is for in vitro diagnostic use.
Brand
Kamiya Biomedical Company, LLC
Lot Codes / Batch Numbers
Catalog number: "KAI-084C", Lot # "559011"
Products Sold
Catalog number: "KAI-084C", Lot # "559011"
Kamiya Biomedical Company, LLC is recalling "K-ASSAY CRP (3) Calibrator E". This product is an IVD calibrator for use on automated clinical ch due to The long-term stability of level 2 (1.0 mg/dL) may not be within the specifications due to a decrease in CRP concentration.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The long-term stability of level 2 (1.0 mg/dL) may not be within the specifications due to a decrease in CRP concentration.
Recommended Action
Per FDA guidance
On May 1, 2009 the recalling firm, Kamiya Biomedical Company contacted the 3 consignees by telephone and notified them of the issue. A follow-up e-mail notification on May 4, 2009 was sent to consignees requesting to return the affected product to Kamiya Biomedical Company for a free replacement kit. After the e-mail notifications, an Urgent Medical Device Recall Notice dated May 13, 2009 was also sent to consignees describing the issue and actions to take for returning opened and unopened affected product to the recalling firm for replacement. Direct questions to the Kamiya Biomedical Company by calling 1- 206-575-8068.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.