Karl Storz Endoscopy Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178 x 35 mm, REF: 11580C; Metal Tray, 275 x 175 x 37 mm, REF: 11580D; Wire Tray f. Cleaning, 290x60x52mm, REF: 39501A1; Wire Tray f. Cleaning, 352 x 125 x 54 mm, REF: 39501A2; Wire Tray f. Cleaning, 487 x 125 x 54 mm, REF: 39501B2; Wire Tray for ENDOCAMELEON, REF: 39501BEC; Tray, with lid, 670x80x52mm, REF: 39501C; Wire Tray for ENDOCAMELEON, long version, REF: 39501CEC; Wire Tray Retromolar Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178 x 35 mm, REF: 11580C; Metal Tray, 275 x 175 x 37 mm, REF: 11580D; Wire Tray f. Cleaning, 290x60x52mm, REF: 39501A1; Wire Tray f. Cleaning, 352 x 125 x 54 mm, REF: 39501A2; Wire Tray f. Cleaning, 487 x 125 x 54 mm, REF: 39501B2; Wire Tray for ENDOCAMELEON, REF: 39501BEC; Tray, with lid, 670x80x52mm, REF: 39501C; Wire Tray for ENDOCAMELEON, long version, REF: 39501CEC; Wire Tray Retromolar
Brand
Karl Storz Endoscopy
Lot Codes / Batch Numbers
REF/UDI-DI: 11580A/ 04048551051801, 11580B/ 04048551051818, 11580C/ 04048551051825, 11580D/ 04048551051832, 39501A1/ 04048551157619, 39501A2/ 04048551157626, 39501B2/ 04048551157657, 39501BEC/ 04048551241400, 39501C/ 04048551157671, 39501CEC/ 04048551254158, 39501F/ 04048551157695, 39501LC2/ 04048551273432, 39501X/ 04048551157756, 39501XK/ 04048551157763, 39501XP/ 04048551157770, 39550A/ 04048551157879. All lots.
Products Sold
REF/UDI-DI: 11580A/ 04048551051801; 11580B/ 04048551051818; 11580C/ 04048551051825; 11580D/ 04048551051832; 39501A1/ 04048551157619; 39501A2/ 04048551157626; 39501B2/ 04048551157657; 39501BEC/ 04048551241400; 39501C/ 04048551157671; 39501CEC/ 04048551254158; 39501F/ 04048551157695; 39501LC2/ 04048551273432; 39501X/ 04048551157756; 39501XK/ 04048551157763; 39501XP/ 04048551157770; 39550A/ 04048551157879. All lots.
Karl Storz Endoscopy is recalling Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, due to Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes information regarding the sterilization modality.
Recommended Action
Per FDA guidance
On 4/6/23, correction notices were distributed to customers who were informed the following: Affected metal trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in pre-vacuum steam sterilization cycles (132C for 4 minutes with a 20-minute dry time). When used in conjunction with an FDA-cleared sterilization wrap, sterility of the enclosed medical device is maintained until used. The system has been validated with devices with working channels greater-than-or-equal-to 0.2 mm and a maximum length of 240 mm. 1) Ensure that all relevant personnel in your organization review this letter, including the new Instructions for Use for the metal trays: following: https://spwebspace.karlstorz.com/sites/MetalTrays/SitePages/Home.aspx 2) If you have previously sold or transferred any of the affected metal trays to a third party, please forward this notice to all such third parties. 3) Complete and return the Acknowledgment Form to karlstorz5983@sedgwick.com Customers correction-related questions can contact karlstorz5983@sedgwick.com or 888-943-5128. Customers with device-related questions can call 800-421-0837
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026