KCI USA, Inc. 500 mL InfoV.A.C. Canisters (with Gel); Part #M8275063/5 (5 canister pack) and Part #M8275063/10 (10 canister pack); products are single use and labeled as STERILE; distributed by Kinetic Concepts, Inc., San Antonio, TX. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
500 mL InfoV.A.C. Canisters (with Gel); Part #M8275063/5 (5 canister pack) and Part #M8275063/10 (10 canister pack); products are single use and labeled as STERILE; distributed by Kinetic Concepts, Inc., San Antonio, TX.
Brand
KCI USA, Inc.
Lot Codes / Batch Numbers
For M8275063/5: 20064392, 20069119, 20069125, 20069126, 20069279, 20069279.18308F, 20075082, and 20079173 For M8275063/10: 20046268, 20046268.18908A, 20046270, 20063933, 20064393, 20064394, 20064395, 20064753, 20064753.14408B, 20067300, 20067300.14408B, 20067301, 20067302, 20067303, 20067304, 20067305, 20069120, 20069121, 20069273, 20069273.14408B, 20069274, 20069275, 20069275.18908A, 20069277, 20069277.18908A, 20069278, 20069278.18908A, 20069280, 20075084, 20075085, 20075086, 20075262, 20075263, 20075264, 20075265, 20075266, 20075268, 20075269, 20079174, 20079175, 20079176, 20079177, 20079178, 20079179, 20079180, 20079181, 20079182, 20079183, 20079184, 20079185, 20079186, 20079187, 20079188, 20089680, 20089681, 20089682, 20089712, 20089713, 20089749, 20089750, 20093372, 20093374, and 20094200
Products Sold
For M8275063/5: 20064392, 20069119, 20069125, 20069126, 20069279, 20069279.18308F, 20075082, and 20079173 For M8275063/10: 20046268, 20046268.18908A, 20046270, 20063933, 20064393, 20064394, 20064395, 20064753, 20064753.14408B, 20067300, 20067300.14408B, 20067301, 20067302, 20067303, 20067304, 20067305, 20069120, 20069121, 20069273, 20069273.14408B; 20069274, 20069275, 20069275.18908A, 20069277, 20069277.18908A, 20069278, 20069278.18908A, 20069280, 20075084, 20075085, 20075086, 20075262, 20075263, 20075264, 20075265, 20075266, 20075268, 20075269, 20079174, 20079175, 20079176, 20079177, 20079178, 20079179, 20079180, 20079181, 20079182, 20079183, 20079184, 20079185, 20079186, 20079187, 20079188, 20089680, 20089681, 20089682, 20089712, 20089713, 20089749, 20089750, 20093372, 20093374, and 20094200
KCI USA, Inc. is recalling 500 mL InfoV.A.C. Canisters (with Gel); Part #M8275063/5 (5 canister pack) and Part #M8275063/10 (10 due to The port on the 500mL InfoV.A.C. Canister that connects to the suction pump tubing was found to be partially or fully occluded with plastic.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The port on the 500mL InfoV.A.C. Canister that connects to the suction pump tubing was found to be partially or fully occluded with plastic.
Recommended Action
Per FDA guidance
Urgent - Medical Device Recall letters were sent within the US on 07/29/08 via UPS next day delivery service and a return receipt was maintained to demonstrate delivery of the recall to all consignees. Additionally, firm is sending representatives into accounts which received more than three sales units of canisters and to all foreign consignee locations to facilitate quarantine and replacement of affected stock. Consignees asked to confirm that adequate flow is being delivered and that the dressing is being drawn down for those patients currently using the InfoV.A.C. Canisters. Also, consignees are to locate and quarantine all packaged 500 ml canisters without the green inspection stickers from within the affected lot numbers in their inventory. A KCI Rep will contact them to help facilitate this process. Consignees are to complete and return the Recall Acknowledgement and Inventory Accounting Form. KCI will replace any affected product at their facility and arrange for the return of any affected product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026