Kentec Medical Inc Reusable Ohmeda Compatible Finger Clip Sensor, Model Number: E403-02, Product Code: PS2225. To monitor oxygen saturation level of patients under doctor's care. Kentec Medical Inc., Irvine, CA 92614 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Reusable Ohmeda Compatible Finger Clip Sensor, Model Number: E403-02, Product Code: PS2225. To monitor oxygen saturation level of patients under doctor's care. Kentec Medical Inc., Irvine, CA 92614
Brand
Kentec Medical Inc
Lot Codes / Batch Numbers
Serial Numbers: YE508/001, YE508/002, YE508/003, YE508/004, YE508/005, YE508/006, YE508/007, YE508/008, YE508/009, YE508/010, ZC502/001, ZC502/002, ZC502/003, ZC502/004, ZC502/005, ZC502/006, ZG512/001, ZG512/002, ZG512/003, ZG512/004, ZG512/005, ZG512/006, ZG512/007, ZG512/008, ZG512/009, ZG512/010, ZG512/011, ZG512/012, ZG512/013, ZG512/014, ZG512/015, PA514/001, PA514/002, PA514/003, PA514/004, PA514/005, PA514/006, PA514/006, PA514/007, PA514/008, PA514/009, PA514/010, PA515/001, PA515/002, PA515/003, PA515/004, PA515/005, PA515/006, PA515/007, PA515/008, PA515/009, PA515/010, PA515/011, PA515/012, PA515/013, PA515/014, PA515/015, PA515/016, PA515/017, PA515/018, PA515/019, PA515/020, PA515/021, PA515/022, PA515/023, PA515/024, PA515/025, PA515/026, PA515/027, PA515/028, PA515/029, PA515/030, PA515/031, PA515/032, PA515/033, PA515/034 and PA515/035
Products Sold
Serial Numbers: YE508/001, YE508/002, YE508/003, YE508/004, YE508/005, YE508/006, YE508/007, YE508/008, YE508/009, YE508/010; ZC502/001, ZC502/002, ZC502/003, ZC502/004, ZC502/005, ZC502/006; ZG512/001, ZG512/002, ZG512/003, ZG512/004, ZG512/005, ZG512/006, ZG512/007, ZG512/008, ZG512/009, ZG512/010, ZG512/011, ZG512/012, ZG512/013, ZG512/014, ZG512/015; PA514/001, PA514/002, PA514/003, PA514/004, PA514/005, PA514/006, PA514/006, PA514/007, PA514/008, PA514/009, PA514/010; PA515/001, PA515/002, PA515/003, PA515/004, PA515/005, PA515/006, PA515/007, PA515/008, PA515/009, PA515/010, PA515/011, PA515/012, PA515/013, PA515/014, PA515/015, PA515/016, PA515/017, PA515/018, PA515/019, PA515/020, PA515/021, PA515/022, PA515/023, PA515/024, PA515/025, PA515/026, PA515/027, PA515/028, PA515/029, PA515/030, PA515/031, PA515/032, PA515/033, PA515/034 and PA515/035
Kentec Medical Inc is recalling Reusable Ohmeda Compatible Finger Clip Sensor, Model Number: E403-02, Product Code: PS2225. To moni due to Incompatible components: The Finger Clip Sensors - with a Serial Number prefix of PA514, PA515, YE508, ZC502 and ZG512 - are missing an additional di. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Incompatible components: The Finger Clip Sensors - with a Serial Number prefix of PA514, PA515, YE508, ZC502 and ZG512 - are missing an additional diode and higher resistor in its existing configuration making them inoperable when used to interface with those Ohmeda Tuffsat monitors previously re-configured by Ohmeda to include a detection circuit for diodes.
Recommended Action
Per FDA guidance
Consignees were notified via telephone on 5/18/07 and an "Important Voluntary Market Withdrawal Notice" letter on 5/23/07. The letter instructed consignees to segregate and return all affected devices. The letter included an "Acknowledgement of Market Withdrawal Notification" form for customer to complete and return to the recalling firm. For additional information, contact 1-800-825-5996.