Klarity Medical Products LLC Klarity Vacuum Bags, Item Nos. R7504-30C, R7504-35BC,R7504-35C, R7644-32.5BC, RC7221-25BC, RC7504-17.5C - Product Usage: To be used by trained medical professionals for the stable support and positioning of patients undergoing external beam radiation therapy treatment in a clinic or hospital setting. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Klarity Vacuum Bags, Item Nos. R7504-30C, R7504-35BC,R7504-35C, R7644-32.5BC, RC7221-25BC, RC7504-17.5C - Product Usage: To be used by trained medical professionals for the stable support and positioning of patients undergoing external beam radiation therapy treatment in a clinic or hospital setting.
Brand
Klarity Medical Products LLC
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Lots R7504-30C: 70621Y and 81008X R7504-35BC: 70415V and 71220V R7504-35C: 81008W R7644-32.5BC: 71228S RC7221-25BC: 70502Y and 70621W RC7504-17.5C: 80411A and 80802W
Klarity Medical Products LLC is recalling Klarity Vacuum Bags, Item Nos. R7504-30C, R7504-35BC,R7504-35C, R7644-32.5BC, RC7221-25BC, RC7504-17 due to Vacuum bags potentially may not hold their vacuum adequately for needed patient treatment times.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Vacuum bags potentially may not hold their vacuum adequately for needed patient treatment times.
Recommended Action
Per FDA guidance
On July 31, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers, informing them that vacuum bags may experience product failure wherein the bags fail to maintain an adequate vacuum over a required period of time. The loss of vacuum can lead to treatment inaccuracy due to shift of patient location. Customers were instructed to: 1. Inspect current inventory to locate any affected product. 2. Remove product from service. 3. Destroy and dispose of product (the firm recommended cutting up and disposing with normal trash). Klarity will replace affected bags or reimburse your cost, as you choose. 4. Complete the customer response form to confirm receipt of notice and actions taken. If any questions, please contact: Sherry Angus, Klarity Quality Manager. Ph. 740-788-8107 ext.119, sherry@klaritymedical.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026