Laboratoires Bodycad, Inc. Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Brand
Laboratoires Bodycad, Inc.
Lot Codes / Batch Numbers
Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.
Products Sold
Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.
Laboratoires Bodycad, Inc. is recalling Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine due to There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data vali. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Recommended Action
Per FDA guidance
The recalling firm issued an Advisory Notice Information dated 2020-06-30 on 7/2/2021 via email which explained the reason for recall and that they would be contacted as soon as possible to proceed with the exchange. On 9/24/2021, a Returned Goods Authorization Form was sent to the consignees so the adapter could be exchanged.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CO, MI, NY