Lang Dental Mfg Co, Inc. Jet Denture Repair Acryulic, fast curing denture repair resin; Lang Dental Manufacturing Co., Inc., Wheeling, IL 60090-0969; product code 1220, 100 gram plastic squeeze bottle and product code 1223 100 gram plastic squeeze bottle, professional package Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Jet Denture Repair Acryulic, fast curing denture repair resin; Lang Dental Manufacturing Co., Inc., Wheeling, IL 60090-0969; product code 1220, 100 gram plastic squeeze bottle and product code 1223 100 gram plastic squeeze bottle, professional package
Brand
Lang Dental Mfg Co, Inc.
Lot Codes / Batch Numbers
product code 1220, box lot 1220-4023, bottle lot 60274005/04AB - clear, product code 1223, box lot 1223-4024, bottle lot 60274005/04AB -clear and box lot 1223-4025, bottle lot 12163001/01AN - meharry
Products Sold
product code 1220, box lot 1220-4023, bottle lot 60274005/04AB - clear; product code 1223, box lot 1223-4024, bottle lot 60274005/04AB -clear and box lot 1223-4025, bottle lot 12163001/01AN - meharry
Lang Dental Mfg Co, Inc. is recalling Jet Denture Repair Acryulic, fast curing denture repair resin; Lang Dental Manufacturing Co., Inc., due to The dental acrylic resin powders are contaminated with small black particles believed to be polyethylene.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The dental acrylic resin powders are contaminated with small black particles believed to be polyethylene.
Recommended Action
Per FDA guidance
Lang Dental sent recall letters dated 6/7/04 to their direct accounts, informing them of the contamination of the acrylic resin powders with the black particles and requesting the return of any existing inventory of the affected lots of product for replacement. Distributors were requested to sub-recall the products from their accounts. Follow-up letters were sent to the accounts on 6/14/04, clarifying the lot numbers affected to include both the lot number on the box and the lot number on the bottom of the squeeze bottles affected.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026