Ion Torrent Dx Control Kit (Life Technologies) – Expiration Extension (2020)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Ion Torrent Dx No Template Control Kit- IVD for detection of DNA and RNA isolated from non-small cell lung cancer (NSCLC) tumor specimens Catalog Number: A32444
Brand
Life Technologies Corporation
Lot Codes / Batch Numbers
the approved shelf life of the vials is also 9 months. Please note, which is also incorrect UDI: (01)10190302006156
Products Sold
Lot Codes: S119_2174740* S119_2108429* P119_2149308 N119_2078115 S218_2069396 V118_1991155 T118_1990947 S217_2008385 S118_1988604 P118_1996942 N118_1974439 S117_1988308 * Mislabeled with a 24-month expiration date. However, the approved shelf life of the vials is also 9 months. Please note, the kit level labelling for the two impacted lots of Ion Torrent Dx No Template Control kit shows a 12-month expiration date, which is also incorrect UDI: (01)10190302006156
Life Technologies Corporation is recalling Ion Torrent Dx No Template Control Kit- IVD for detection of DNA and RNA isolated from non-small cel due to Extended Expiration Date prior to clearance. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Extended Expiration Date prior to clearance
Recommended Action
Per FDA guidance
Thermo Fisher issued notification on 3/16/20 via email followed by hard copy on 3/18/20, stating reason for recall, health risk and action to take: Review the expiration date of the products referenced in the table included in the Customer Notification and take the following actions: "Notify all Oncomine Dx Target Test users in your facility of this issue. "Review the expiration date of the products referenced in the table above and using Attachment 1 of this letter, take the following actions. Attachment 1 provides the corrected expiration dates per lot. For products that are past the 9-month corrected expiration date per Attachment 1: Discontinue use of the product, record the kit lot numbers and quantities impacted on the attached Customer Response Sheet, and email the completed response sheet to gabriella.armin@thermofisher.com. For products that are still within the 9-month corrected expiration date per Attachment 1: Correct the expiration dating on the product labelling within your inventory by reviewing Attachment 1 to determine the correct expiration date per kit lot. Strike through the incorrect expiration date and add the new one using permanent ink or amend the label by adding an additional label correcting the expiration date on the labelling. The updated labelled product can continue to be used until the 9-month expiration date, after which any remaining product should be discarded. Note: If affected product was used past the 9-month corrected expiration date per Attachment 1, but before the 12-month expiration date on the kit labels, there is no anticipated risk of generating inaccurate/incorrect results because the system will inform the user if the assay did not meet run passing criteria and test results wont be generated. Therefore, no action on your part is required. If you have any technical questions or concerns, please contact Technical Support at [NA --800-955-6288 option 4 or techsupport@thermofisher.com]
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026