Magellan Diagnostics, Inc. LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
Brand
Magellan Diagnostics, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
UDI-DI :(01)00850355006000 Lot Numbers: 2231 2231-10 2303M-11 2303M-12 2303M-13 2303M-15 2305M-01 2305M-02 2305M-03 2305M-04 2305M-05 2306M-01 2306M-02 2306M-03 2306M-04 2306M-05 2306M-06 2307M 2307M-04 2307M-05 2307M-06 2307M-07 2307M-10 2307M-11 2308M-01 2308M-02 2308M-03 2308M-07 2309M 2310M-01 2310M-03 2310M-04 2310M-05 2310M-06 2310M-08 2311M-02 2311M-03 2311M-05 2311M-09 2311M-10 2312M 2312M-01 2312M-04 2312M-05 2312M-07 2313M 2313M-01 2313M-05 2313M-07 2314M-06 2314M-07 2314M-08 2314M-09 2314M-12 2314M-13 2315M-02 2315M-03 2315M-04 2315M-05 2315M-06 2315M-07 2315M-08 2315M-13 2316M 2316M-03 2316M-04 2316M-05 2316M-06 2316M-11 2316M-12 2317M 2317M-01 2318M 2320M-06 7308M
Magellan Diagnostics, Inc. is recalling LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitati due to During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimens. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
Recommended Action
Per FDA guidance
Magellan notified distributors/end users. Distributors initially contacted 19 September 2023 and told they would receive additional contacts from Magellan related to the recall and requests for a formal response. All distributors contacted by phone and provided the Recall Response forms included in the Recall Notification Letters. Hard copy recall notification letters issued via mail on Sept. 22, 2023. Distributors requested to provide customer lists and notifications will be carried out to the user level. Distributors will be permitted to continue distribution of impacted product, and users will be instructed to continue to use impacted product; however, replacement product will be provided upon request or if a consignee complains of PATFP or COOR-Hi results. The recall letter states reason for recall, health risk and action to take: Review current inventory and identify any remaining unexpired stock impacted by this recall: MAGELLAN WILL REPLACE IMPACTED PRODUCT UPON REQUEST; HOWEVER,YOU MAY CONTINUE TO USE ANY REMAINING UNEXPIRED STOCK. Continue to monitor impacted product for Controls Out of Range-High (COOR-Hi) results. If after following instructions in the LeadCare II Users Guide the control value is still out of range, please contact Magellan Technical Support. If you observe an elevated blood lead result on a patient, the test results should be shared with the patients physician for interpretation and to determine when retesting and follow-up care are necessary. If you identify a PATFP or a COOR-Hi result, you can contact Magellan Technical Support at 1-800-275-0102 to obtain a FedEx label to return any product to Magellan Diagnostics, Inc. and receive replacement product. Product will be replaced based on availability, and on a first-come first-served basis. NO ADDITIONAL ACTION IS REQUIRED: It is the recommendation of the Centers for Disease Control and Prevention (CDC) that if an initial blood lead screening is above the Blood Lead Reference Valu
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.