Maquet Cardiovascular, LLC Fusion Bioline Vascular Graft, Part number M00201503046B0 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Fusion Bioline Vascular Graft, Part number M00201503046B0
Brand
Maquet Cardiovascular, LLC
Lot Codes / Batch Numbers
UDI-DI 00384409005942 Lot 25162546
Products Sold
UDI-DI 00384409005942 Lot 25162546
Maquet Cardiovascular, LLC is recalling Fusion Bioline Vascular Graft, Part number M00201503046B0 due to One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distribut. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.
Recommended Action
Per FDA guidance
On January 30, 2023, the firm notified customers via Urgent Medical Device Removal letters. Customers are asked to examine your inventory immediately for lot 25162546. Please note, the Lot number appears on the device shelf boxes only. If you have the affected lot, please stop using and remove the device from areas of use. You are entitled to a replacement at no cost to your facility. You will receive replacement upon your acknowledgement that you have product for return. Distributors who have shipped this product should notify their customers. Please contact Getinge Customer Service at 888-943-8872 between the hours of 8:00 a.m. and 6:00 p.m. Eastern Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026