MAQUET Inc. Foot switch, Part Number 1009.81. Maquet GmbH & Co. KG, 76437 Rastatt, Germany. The foot switches in the 1009.8 1 series are accessory components which can be connected to certain MAQUET operating tables. Several of the table top adjustment motions (depending on the table and foot switch versions in use) can then be executed by the foot. The foot switch is equipped with rocker switches. The switching elements are used to initiate the table top functions. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Foot switch, Part Number 1009.81. Maquet GmbH & Co. KG, 76437 Rastatt, Germany. The foot switches in the 1009.8 1 series are accessory components which can be connected to certain MAQUET operating tables. Several of the table top adjustment motions (depending on the table and foot switch versions in use) can then be executed by the foot. The foot switch is equipped with rocker switches. The switching elements are used to initiate the table top functions.
Brand
MAQUET Inc.
Lot Codes / Batch Numbers
Product manufactued between 8/2004 and 9/2008 may be affected: 1009.81A0, 1009.81A1, 1009.81A2, 1009.81A3, 1009.81B0, 1009.81B1, 1009.81B2, 1009.81C0, 1009.81C1, 1009.81D0, 1009.81D1, 1009.81D2, 1009.81D3, 1009.81D4, 1009.81E0, 1009.81E1, 1009.81E2, 1009.81E3, 1009.81F0, 1009.81F1, 1009.81F2, 1009.81F3, 1009.81G0, 1009.81G1 and 1009.81G2.
Products Sold
Product manufactued between 8/2004 and 9/2008 may be affected: 1009.81A0, 1009.81A1, 1009.81A2, 1009.81A3, 1009.81B0, 1009.81B1, 1009.81B2, 1009.81C0, 1009.81C1, 1009.81D0, 1009.81D1, 1009.81D2, 1009.81D3, 1009.81D4, 1009.81E0, 1009.81E1, 1009.81E2, 1009.81E3, 1009.81F0, 1009.81F1, 1009.81F2, 1009.81F3, 1009.81G0, 1009.81G1 and 1009.81G2.
MAQUET Inc. is recalling Foot switch, Part Number 1009.81. Maquet GmbH & Co. KG, 76437 Rastatt, Germany. The foot switches due to A bearing point of the pedal of the foot switch may come loose. With continued use of the foot switch, one or both springs beneath the pedal can fall. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A bearing point of the pedal of the foot switch may come loose. With continued use of the foot switch, one or both springs beneath the pedal can fall out of their mounting, and it is possible the pedal comes to rest on the switch causing an unintended motion of the operating table during patient use.
Recommended Action
Per FDA guidance
An Urgent Device Correction letter (with an enclosed Customer Response form) was sent via Federal Express on Monday, December 8, 2008 to all customers who had received tables. Customers were notified in the letter of the problem and actions to be taken including not to use any affected foot pedal for safety reasons. In addition, customers were instructed to complete and return the Customer Response Form via fax (732-667-1903) to Maquet Inc. Please call USA Maquet, Inc. at toll free number 1-888-627-8383 #1 (then select the option for Surgical Workplace products) for questions.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026