Maquet Inc. Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS operating table system. Used to mount head positioning accessories. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS operating table system. Used to mount head positioning accessories.
Brand
Maquet Inc.
Lot Codes / Batch Numbers
No serial or lot numbers, just a catalogue number.
Products Sold
No serial or lot numbers, just a catalogue number.
Maquet Inc. is recalling Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGN due to Product monitoring has determined that the connection bracket shows potential mechanical play at the OR table interface for some procedures. This is d. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Product monitoring has determined that the connection bracket shows potential mechanical play at the OR table interface for some procedures. This is due to the hole in the lock washer, which is responsible for locking the accessory to the OR table, being too large.
Recommended Action
Per FDA guidance
Urgent Device Correction Letters were sent by Federal Express to affected US hospitals on October 22, 2009. The letter stated the technical description, potential hazard, and the recommended precaution. The letter asked customers to immediately verify whether their Connection Bracket is affected thru the FaxBack provided. Questions or need for additional information can be directed to customers' local MAQUET representative.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026