Maquet Medical Systems USA BEQ-HMOD70000-USA/QUADROX-iD, Oxygenator used during cardiopulmonary bypass Item no. 701067859 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BEQ-HMOD70000-USA/QUADROX-iD, Oxygenator used during cardiopulmonary bypass Item no. 701067859
Brand
Maquet Medical Systems USA
Lot Codes / Batch Numbers
UDI-DI: 04058863019024 All lots
Products Sold
UDI-DI: 04058863019024 All lots
Maquet Medical Systems USA is recalling BEQ-HMOD70000-USA/QUADROX-iD, Oxygenator used during cardiopulmonary bypass Item no. 701067859 due to 1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory s sterile pouch. 2. Product packaging may exhib. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory s sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.
Recommended Action
Per FDA guidance
On May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 between the hours of 6:00 AM and 5:00 PM Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. Distributors who have shipped affected products to customers should forward the notification to customers' attention for appropriate action. Maquet Cardiopulmonary GmbH is working with all possible urgency on the redesign and revalidation of a packaging solution, which is anticipated to be available by Q4 2023.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026