Medefil Incorporated Medefil Heparin I.V. Flush Syringe 10 units/mL; a polypropylene luer locking syringe containing Heparin Lock Flush Solution, USP, a sterile non-pyrogenic, isotonic injection; each mL in a Heparin I.V. Flush Syringe contains 10 units of heparin sodium derived from porcine intestinal mucosa and 9 mg of sodium chloride in water for injection; Medefil, Inc., Glendale Heights, IL 60139 The pre-filled syringes were packed 60 syringes per dispenser box, 16 boxes per shipping carton in the following sy Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medefil Heparin I.V. Flush Syringe 10 units/mL; a polypropylene luer locking syringe containing Heparin Lock Flush Solution, USP, a sterile non-pyrogenic, isotonic injection; each mL in a Heparin I.V. Flush Syringe contains 10 units of heparin sodium derived from porcine intestinal mucosa and 9 mg of sodium chloride in water for injection; Medefil, Inc., Glendale Heights, IL 60139 The pre-filled syringes were packed 60 syringes per dispenser box, 16 boxes per shipping carton in the following sy
Brand
Medefil Incorporated
Lot Codes / Batch Numbers
catalog #MIH-2235: lot H07325, exp. date 11/09, H08113, exp. date 12/09, catalog #MIH-2233: lot H08124, exp. date 01/10
Products Sold
catalog #MIH-2235: lot H07325, exp. date 11/09, H08113, exp. date 12/09; catalog #MIH-2233: lot H08124, exp. date 01/10
Medefil Incorporated is recalling Medefil Heparin I.V. Flush Syringe 10 units/mL; a polypropylene luer locking syringe containing Hepa due to The heparin lock flush solution was manufactured from contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API). The FDA has received re. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The heparin lock flush solution was manufactured from contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API). The FDA has received reports of serious injuries and/or deaths in patients administered finished heparin injectable products manufactured from heparin API containing this contaminant.
Recommended Action
Per FDA guidance
Medefil telephoned the direct accounts on 3/20/08, e-mailed them a copy of the 3/20/08 recall letter on 3/21/08, and sent the accounts a hard copy of the recall letter on 3/24/08. The accounts were informed of Medefil's recall of 19 lots of Heparin I.V. Flush Syringes produced from the active pharmaceutical ingredient (API) Heparin Sodium, USP, lot 1035-0773, that has been recalled by the manufacturer, Scientific Protein Laboratories, due to the presence of a contaminant in the recalled lot of API. The accounts were instructed to stop use of the 19 lots of Heparin I.V. Flush Syringes immediately, place them under quarantine, and contact all customers to whom they further distributed the product, telling them to stop use of the product and return the product to Medefil through their supplier. The accounts were requested to complete and return to Medefil via fax the enclosed response form, indicating the quantity on hand of each recalled lot. Any questions were directed to Pradeep Aggarwal, President at 630-682-4600.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026