Medikmark, Incorporated MediChoice Skin Staple Remover, reorder no. SSR 1001; a sterile Rx convenience kit containing a skin staple remover, a PVP prep pad and a 3 x 3 gauze sponge; 50 kits per case; Made for Owens & Minor Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
MediChoice Skin Staple Remover, reorder no. SSR 1001; a sterile Rx convenience kit containing a skin staple remover, a PVP prep pad and a 3 x 3 gauze sponge; 50 kits per case; Made for Owens & Minor
Brand
Medikmark, Incorporated
Lot Codes / Batch Numbers
reorder no. SSR 1001: lots 305097, 306073, 306212, 312109, 401003, 101133, 403154, 404040, 405003, 406021
Products Sold
reorder no. SSR 1001: lots 305097, 306073, 306212, 312109, 401003, 101133, 403154, 404040, 405003, 406021
Medikmark, Incorporated is recalling MediChoice Skin Staple Remover, reorder no. SSR 1001; a sterile Rx convenience kit containing a skin due to The tip of the staple remover may puncture a hole in the wall of the tray, compromising the sterility of the kit.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The tip of the staple remover may puncture a hole in the wall of the tray, compromising the sterility of the kit.
Recommended Action
Per FDA guidance
Recall notifications were sent by fax and follow-up registered mail to the direct account distributors and users on 7/12-15/04. The accounts were informed of the potential non-sterility of the product and were instructed to immediately discontinue sale/use of the products and place all inventory in quarantine pending return to Medikmark. Distributors were instructed to sub-recall the kits from their customers. Any questions were directed to 1-847-596-7723.