Medline Industries Inc MEDLINE ELECTRIC SURGICAL CLIPPER AND BASE, REF DYND70800 Product Usage: Medline Surgical Clipper is intended for the removal of body hair from the patient as required prior to surgery or other minor procedures. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
MEDLINE ELECTRIC SURGICAL CLIPPER AND BASE, REF DYND70800 Product Usage: Medline Surgical Clipper is intended for the removal of body hair from the patient as required prior to surgery or other minor procedures.
Brand
Medline Industries Inc
Lot Codes / Batch Numbers
14G1
Products Sold
14G1
Medline Industries Inc is recalling MEDLINE ELECTRIC SURGICAL CLIPPER AND BASE, REF DYND70800 Product Usage: Medline Surgical Clippe due to The recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a charger base malfunction.
Recommended Action
Per FDA guidance
Medline Industries initiated the recall on 09/06/2019 by letter. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Immediately check your stock for the affected item number and lot number listed on the destruction form. Quarantine all affected product. Destroy all product belonging to the affected lot numbers listed on the destruction form. 2. Please return the completed enclosed destruction form listing the quantity of affected product destroyed. Even if you do not have any affected product please complete and return the form. For questions contact 866-359-1704.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026