Medline Industries Inc Medline Latex Free Labor Kit & Postpartum CDS-LF; a single patient prescription procedure pack and disposables, including a Gerber Nuk Pacifier; Made in USA by Medline Industries, Inc., Mundelein, IL 60060 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medline Latex Free Labor Kit & Postpartum CDS-LF; a single patient prescription procedure pack and disposables, including a Gerber Nuk Pacifier; Made in USA by Medline Industries, Inc., Mundelein, IL 60060
Brand
Medline Industries Inc
Lot Codes / Batch Numbers
LABOR KIT & POSTPARTUM CDS-LF reorder # CDS830074D, lots 05HB0427, 05IB0654, 05IB1857, 05KB0229, 06AB1244, 06BB0159, 06CB0593, 06CB1926, 06CB2197, 06FB0015, 06FB0476, 06GB0916, 06HB0704, 06HB2791, 06IB1312, 06IB1922, 06IB2566, 06KB0020, 06KB1042, 06KB2322, 06LB2102, 07AB0437, 07BB2798, 05AD3274, 05BD4131, 05CD3200, 05FD2631, reorder #CDS830074F, lots 07DB1578, 07GB1509, 07IB0792, 07JB0627, 07JB1829, 07LB0150, 07LB0508, 08BB0151
Products Sold
LABOR KIT & POSTPARTUM CDS-LF reorder # CDS830074D, lots 05HB0427, 05IB0654, 05IB1857, 05KB0229, 06AB1244, 06BB0159, 06CB0593, 06CB1926, 06CB2197, 06FB0015, 06FB0476, 06GB0916, 06HB0704, 06HB2791, 06IB1312, 06IB1922, 06IB2566, 06KB0020, 06KB1042, 06KB2322, 06LB2102, 07AB0437, 07BB2798, 05AD3274, 05BD4131, 05CD3200, 05FD2631; reorder #CDS830074F, lots 07DB1578, 07GB1509, 07IB0792, 07JB0627, 07JB1829, 07LB0150, 07LB0508, 08BB0151;
Medline Industries Inc is recalling Medline Latex Free Labor Kit & Postpartum CDS-LF; a single patient prescription procedure pack and d due to The latex free labor & delivery packs contain a latex Nuk pacifier. The pacifier is packaged separately and offers no risk of exposure of latex parti. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The latex free labor & delivery packs contain a latex Nuk pacifier. The pacifier is packaged separately and offers no risk of exposure of latex particles to other components of the pack.
Recommended Action
Per FDA guidance
Recall letters dated 8/4/08 were sent to the direct accounts on the same date, informing them that the labor and delivery packs were labeled Latex Free, but contained a Latex Nuk pacifier. The letters were customized for each account, listing the specific affected reorder numbers and lot numbers shipped to each account. The accounts were requested to immediately check their stocks for the affected lot numbers and quarantine the affected product found. The accounts were given three options: A) Attach the enclosed stickers to the packs so that the user is aware of the latex Nuk pacifier, B) Call Medline to receive replacement non-latex pacifiers, but also place the stickers on the packs, C) Return the packs to Medline for replacements. The accounts were requested to complete the enclosed response form, indicating which option the account selected and whether they needed any additional stickers, and fax the form back to Medline at 847-643-4482 by 8/18/08. Any questions were directed to Kathy Dunne at 1-866-359-1704.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026