Medline Industries Inc Medline Strider Maxi 3 Scooter; a three wheeled battery operated scooter; manufactured by KYMCO, Kaohsiung, Taiwan for Medline Industries, Inc., One Medline Place, Mundelein, IL 60060 USA; model MDS807600 - red and MDS807600B - blue Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medline Strider Maxi 3 Scooter; a three wheeled battery operated scooter; manufactured by KYMCO, Kaohsiung, Taiwan for Medline Industries, Inc., One Medline Place, Mundelein, IL 60060 USA; model MDS807600 - red and MDS807600B - blue
Brand
Medline Industries Inc
Lot Codes / Batch Numbers
model MDS807600 - red and model MDS807600B - blue, all serial numbers
Products Sold
model MDS807600 - red and model MDS807600B - blue, all serial numbers
Medline Industries Inc is recalling Medline Strider Maxi 3 Scooter; a three wheeled battery operated scooter; manufactured by KYMCO, Kao due to Under conditions of heavy use, signs of overheating of the main battery cables have been observed.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Under conditions of heavy use, signs of overheating of the main battery cables have been observed.
Recommended Action
Per FDA guidance
Medline sent 'Voluntary Field Correction' letters dated 7/20/06 to the direct accounts on the same date, informing them that the Strider Maxi 3 and Maxi 4 scooters have main battery cables that may overheat under conditions of heavy use, and that Medline will provide an upgrade kit to replace the battery cables. Any questions were directed to 800-289-9793, press 1, then press 3 and ask to transfer to 4674 (8:00 AM to 5:00 PM CST). Follow-up letters were sent to the distributors and end-users on 10/31/06. The end users were informed of the problem with the scooters, and were requested to call Medline at 800-289-9793, press 1, then press 3 and ask to transfer to 4674 (8:00 AM to 5:00 PM CST) to have the scooter upgraded. The end users were requested to inform the service technician that hey have one of the affected Maxi Scooters, and provide him with their full name, phone number and street address, as well as the name, address and/or phone number of teh store where the Medline Scooter was purchased. Medline will arrange with the dealership or retail store to upgrade the scooter at no cost to the end user.