MEDLINE INDUSTRIES, LP Northfield Centurion Sterile CirClamp with 1.45cm Bell Reprocessed, Reorder CR3145K, 1 per package, 12 per case. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Centurion Sterile CirClamp with 1.45cm Bell Reprocessed, Reorder CR3145K, 1 per package, 12 per case.
Brand
MEDLINE INDUSTRIES, LP Northfield
Lot Codes / Batch Numbers
Lot #2022032501, exp. 2026-12-31, UDI-DI - (01)10653160312956
Products Sold
Lot #2022032501, exp. 2026-12-31, UDI-DI - (01)10653160312956
MEDLINE INDUSTRIES, LP Northfield is recalling Centurion Sterile CirClamp with 1.45cm Bell Reprocessed, Reorder CR3145K, 1 per package, 12 per case due to The kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size.
Recommended Action
Per FDA guidance
The recalling firm issued letters dated 5/23/2023 via email and first class mail on 5/23/2023 to its consignees, which identified incorrect information regarding the affected product. It had stated the label indicates the kit contains a 1.3 cm bell and insert; however the kit actually contains a 1.45 cm bell and insert. The affected kits were to be destroyed for credit and none were to be returned to the recalling firm. The consignee was to check their stock for the affected item number and lot numbers listed on the recalling firm's recall portal and destroy the affected product. A link was provided to complete the response form. If the consignee is a distributor or has resold or transferred the product to another company or individual, they are to notify them of this recall communication. The recalling firm issued a letter flagged as an Amended Letter dated 9/1/2023 via email and first class mail that alerted the consignee to the incorrect product information contained in the initial 5/23/2023 letter and the Amended Letter provided the correct product information, that being the product is supposed to include a 1.45 cm base and 1.45cm bell, however, the kit incorrectly contains CirClamp subassembly 345CRSAK which includes a 1.6cm base and 1.6cm bell. The letter instructs that affected kits can be destroyed for credit and no kits are to be returned to Medline. The consignee is instructed to immediately check their stock for the item number and lot number listed in the letter and on the destruction form, quarantine the affected product, and destroy it. The enclosed destruction form listing the quantity of affected product destroyed is to be returned. If the product has been transferred to another individual, department or location , the consignee is to notify them of this recall communication.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026