Medline Luer Lock Syringe – Volume Accuracy (2024)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
MEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101010
Brand
MEDLINE INDUSTRIES, LP Northfield
Lot Codes / Batch Numbers
UDI/DI 10884389596319, All lots in the format: 897YYMMXXXX
Products Sold
UDI/DI 10884389596319, All lots in the format: 897YYMMXXXX
MEDLINE INDUSTRIES, LP Northfield is recalling MEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101010 due to It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range o. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
Recommended Action
Per FDA guidance
Medline Industries, LP, issued a recall notice to its consignees on 04/22/2024 via USPS first class mail. The notice explained the problem with the device, potential risk, and requested that the affected product be destroyed.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026