Medtronic Gastroenterology Urology Medtronic Zinetics 24ME Multi-Use External Reference pH Catheter. For Gastroesophageal pH Measurements. Catalog number 9012P2121. Distributed by Medtronic, Inc., 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604, USA. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medtronic Zinetics 24ME Multi-Use External Reference pH Catheter. For Gastroesophageal pH Measurements. Catalog number 9012P2121. Distributed by Medtronic, Inc., 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604, USA.
Brand
Medtronic Gastroenterology Urology
Lot Codes / Batch Numbers
Lot 0001491
Products Sold
Lot 0001491
Medtronic Gastroenterology Urology is recalling Medtronic Zinetics 24ME Multi-Use External Reference pH Catheter. For Gastroesophageal pH Measureme due to Lot 0001491 has a manufacturing defect in the sensor spacing. Sensors have been placed 5 cm apart instead of 15 cm apart. Medtronic has determined t. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Lot 0001491 has a manufacturing defect in the sensor spacing. Sensors have been placed 5 cm apart instead of 15 cm apart. Medtronic has determined that patient safety is not compromised. Clinicians however should review and reassess the results of any patient studies for which catheters from lot 0001491 were used.
Recommended Action
Per FDA guidance
An Urgent Recall letter was issued to all affected customers 1/25/2006. This letter described the issue and listed what actions customers were to take to return affected product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
MI, PA, TX