Medtronic, Inc. 6F Z2 MP1 110CM CATHETER Z26MP1A Z2 6F 110CM MP1 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peripheral vascular system. Product Code: Z26MP1A Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
6F Z2 MP1 110CM CATHETER Z26MP1A Z2 6F 110CM MP1 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peripheral vascular system. Product Code: Z26MP1A
Brand
Medtronic, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Lot Numbers: 0000007859 0000087103 0000087253 0000091588 0000092982 0000103702 0000142221 0000148507 0000182815
Medtronic, Inc. is recalling 6F Z2 MP1 110CM CATHETER Z26MP1A Z2 6F 110CM MP1 Coronary Guiding catheter designed to provide a pa due to Potential for non-sterility due to loss of package integrity. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for non-sterility due to loss of package integrity
Recommended Action
Per FDA guidance
Medtronic notified accounts and sales representatives on 4/05/06 advising the recalled product be segregated and held for return to the sales rep. and documented on the reconciliation form.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026