Medtronic Inc. Cardiac Rhythm Managment Medtronic InSync Marquis, model 7277 Dual Chamber Implantable Cardioverter Defibrillator (ICD) with Cardiac Resynchronization Therapy devices are tiered therapy implantable arrhythmia control devices, which provide automatic detection of ventricular arrhythmias and delivery of user-selected therapies for the detected arrhythmia. Therapies are automatically delivered following detection or can be initiated via programming by the physician. These ICDs are multi-programmable and have the capabili Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medtronic InSync Marquis, model 7277 Dual Chamber Implantable Cardioverter Defibrillator (ICD) with Cardiac Resynchronization Therapy devices are tiered therapy implantable arrhythmia control devices, which provide automatic detection of ventricular arrhythmias and delivery of user-selected therapies for the detected arrhythmia. Therapies are automatically delivered following detection or can be initiated via programming by the physician. These ICDs are multi-programmable and have the capabili
Brand
Medtronic Inc. Cardiac Rhythm Managment
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Model 7277 InSync Marquis DR (serial numbers listed include all affected devices but may include some unaffected devices. Patients are recommended to check with their doctor for exact serial numbers affected.) H-Series: PLT100002H - PLT100583H. S-Series: PLT600001S - PLT611901S.
Medtronic Inc. Cardiac Rhythm Managment is recalling Medtronic InSync Marquis, model 7277 Dual Chamber Implantable Cardioverter Defibrillator (ICD) with due to Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs, depletion can take place within a few hours to a few days, after which there is a complete loss of device function.
Recommended Action
Per FDA guidance
Medtronic Sales Representatives contacted affected physicians starting 02/10/05 to deliver an Important Patient Management Information. The letter recommended patient managment options including quarterly follow-ups, programming Patient Alert to 'On-High', provide magnet to patients to check device status and inform patients that should they expereicnce warmth in the area surrounding the ICD to seek prompt care.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.