Medtronic Inc, Neurological & Spinal Division Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424 Itrel II Neurostimulator (DBS applications only); Model 7426 Soletra Neurostimulator; Model 7428 Kinetra Neurostimulator; Models 3387, 3389 DBS Leads; and Models 7482,7495 DBS Extensions. Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA. The Medtronic Activa System is an implantable, multiprogrammable system that delivers electrical stimulation to selected areas o Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424 Itrel II Neurostimulator (DBS applications only); Model 7426 Soletra Neurostimulator; Model 7428 Kinetra Neurostimulator; Models 3387, 3389 DBS Leads; and Models 7482,7495 DBS Extensions. Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA. The Medtronic Activa System is an implantable, multiprogrammable system that delivers electrical stimulation to selected areas o
Brand
Medtronic Inc, Neurological & Spinal Division
Lot Codes / Batch Numbers
All serial numbers and Lot codes are affected.
Products Sold
All serial numbers and Lot codes are affected.
Medtronic Inc, Neurological & Spinal Division is recalling Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Mod due to A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Me. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Medtronic's instructions have until now recommended a use of a displayed MRI head specific absorption rate (SAR) of not more than 0.4 W/kg. The recommended maximum displayed head SAR is now 0.1 W/kg.
Recommended Action
Per FDA guidance
The firm''s recall consists of providing revised instructions for the Activa DBS. The revision modifies the recommended SAR to use during MRI of patients who have an Activa DBS. Use of the Activa DBS is otherwise unaffected by this recall. A Dear Healthcare Provider letter and MRI Radiologist letter were sent on 11/30/05 and 12/05/05 to provide this information.