Medtronic MiniMed Medtronic MiniMed Paradigm Insulin Infusion Pumps for the following Paradigm models: MMT-511, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medtronic MiniMed Paradigm Insulin Infusion Pumps for the following Paradigm models: MMT-511, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722
Brand
Medtronic MiniMed
Lot Codes / Batch Numbers
"All serial numbers" for the following Paradigm models: MMT-51 1, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722
Products Sold
"All serial numbers" for the following Paradigm models: MMT-51 1, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722
Medtronic MiniMed is recalling Medtronic MiniMed Paradigm Insulin Infusion Pumps for the following Paradigm models: MMT-511, MMT-51 due to Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed P. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemi
Recommended Action
Per FDA guidance
Firm mailed Medical Device Safety letters on April 24, 2007, to healthcare professionals, existing pump users and MRI facilities reiterating the existing warning in the pump User Guide to avoid exposing the pump to MRI (or similar high strength electromagnetic fields) and strengthening previous warning by specifically mentioning the potential for over-delivery and severe hypoglycemia. Firm is also including an insert with this information with any Paradigm infusion pumps shipped to new customers.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
PR