Medtronic Neurological Medtronic Kinetra Dual Program Neurostimulator for Deep Brain Stimulation, model 7428. Rx only. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. The dual program model 7428 Kinetra Neurostimulator generates electrical signals that are transmitted to the brain. These signals are delivered from the neurostimulator to the brain via DBS extension or DBS leads. The neurostimulator consists of electronic circuitry and a battery, which are hermetically sealed in a titanium Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medtronic Kinetra Dual Program Neurostimulator for Deep Brain Stimulation, model 7428. Rx only. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. The dual program model 7428 Kinetra Neurostimulator generates electrical signals that are transmitted to the brain. These signals are delivered from the neurostimulator to the brain via DBS extension or DBS leads. The neurostimulator consists of electronic circuitry and a battery, which are hermetically sealed in a titanium
Brand
Medtronic Neurological
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
serial numbers in the following ranges: ¿ NFD100006 - NFD100325H ¿ NFD620017S - NFD625737S ¿ NFD200201K - NFD201303K
Medtronic Neurological is recalling Medtronic Kinetra Dual Program Neurostimulator for Deep Brain Stimulation, model 7428. Rx only. Me due to A subset of Kinetra implantable neurostimulators may experience a failure of wire connections between the electronic hybrid circuit and battery which . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A subset of Kinetra implantable neurostimulators may experience a failure of wire connections between the electronic hybrid circuit and battery which may lead to sudden cessation of therapy. Sudden cessation of therapy can result in the immediate return or worsening of underlying symptoms due to the progression of the disease state.
Recommended Action
Per FDA guidance
Letter dated July 2006. Recommendations were made to Health Care Providers of the following: 1.Consider the potential for this anomaly if your patient present with a device in a POR state. 2. Consider informing your patients of this risk and advise them to seek attention immediately if they experience a return of pre-implant or more advanced symptoms. 3. Determine whether device replacement is warranted based upon consultation with your patient, review of the patient''s medical history and consideration of the relative risks of abrupt cessation of stimulation versus the risk of a device replacement surgical procedure. 4. Report any malfunction, removal and replacement to Medtronic and FDA suing the medical device reporting system. 5. Acknowledge this communication using the reply form enclosed.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution