Medtronic Neurosurgery Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DU Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DU
Brand
Medtronic Neurosurgery
Lot Codes / Batch Numbers
REF/Expiration Date On or Prior/UDI-DI/Lots: 61100/March 31, 2025/00643169063808/2006039, 2007020, 2009028, 2010001, 2010016, 2102001, 2102020, 2103001, 2105028, 2106001, 2108011, 2108017, 2109022, 2110009, 2110030, 2111023, 2111045, 2112026, 2203020, 2204016, 2204035, 2207014, 2207046, 61105/February 28, 2025/00643169063815/2009060, 2011017, 2102011, 2102027, 2102047, 2103002, 2104042, 2105029, 2108001, 2108012, 2108016, 2109023, 2110010, 2111021, 2111024, 2111046, 2112027, 2202034, 2203034, 2204027, 2206008, 2207015, 2207033, 61106/December 31, 2024/00643169063822/2010017, 2103003, 2105030, 2105038, 2108013, 2108018, 2109004, 2110011, 2110031, 2111025, 2202035, 220302, 61110/March 31, 2025/00643169063839/2006039, 2010018, 2101014, 2102002, 2102028, 2103004, 2105031, 2108019, 2109024, 2110012, 2111026, 2201005, 2202036, 2204025, 2204040, 61111/April 30, 2025/00643169063846/2008039, 2010019, 2102021, 2103005, 2105032, 2106004, 2108014, 2110013, 2111027, 2202037, 2203022, 2203028, 2207016, 62100/January 31, 2025/00643169063853/2007021, 2008040, 2009026, 2009061, 2010014, 2011018, 2012001, 2101015, 2101020, 2102012, 2102022, 2102048, 2103019, 2104043, 2105033, 2106005, 2108020, 2109001, 2109020, 2109028, 2110014, 2110032, 2110033, 2111028, 2111035, 2111037, 2112001, 2112002, 2202038, 2203023, 2204001, 2204028, 2206013, 2206027, 2207017, 2207048, 2208009, 62105/January 31, 2025/00643169063860/2008005, 2008041, 2009029, 2010015, 2010020, 2012002, 2101016, 2102003, 2102013, 2102029, 2103006, 2103036, 2104044, 2105036, 2106035, 2108003, 2108015, 2109002, 2109003, 2110003, 2110015, 2110016, 2111022, 2111033, 2111047, 2201006, 2202014, 2202025, 2203015, 2203025, 2203031, 2204023, 2204036, 2204041, 2206014, 2206028, 2207018, 2207044, 62106/December 31, 2024/00643169063877/2008042, 2010021, 2101017, 2102023, 2103007, 2104045, 2105039, 2106007, 2108004, 2108021, 2109005, 2110017, 2110034, 2111029, 2202039, 2207019, 2207051, 62110/March 31, 2025/00643169063884/2006041, 2007022, 2008006, 2008043, 2010022, 2101009, 2101018, 2102026, 2102049, 2103020, 2104046, 2105001, 2105040, 2108005, 2108022, 2110004, 2110018, 2111036, 2201007, 2202015, 2202031, 2204017, 2204042, 2205031, 2205032, 2206007, 2207020, 2207049, 62111/May 31, 2025/00643169063891/2010023, 2012003, 2101019, 2102024, 2103008, 2105034, 2106009, 2108023, 2109006, 2110019, 2110035, 2111030, 2111034, 2202040, 2203024, 2204002, 2207021, 2207037, 2208027, 62158/April 30, 2025/00643169063907/2011019, 2103009, 2105035, 2110020, 2111031, 2202041, 2203033, 2204029, 2207006
Products Sold
REF/Expiration Date On or Prior/UDI-DI/Lots: 61100/March 31, 2025/00643169063808/2006039, 2007020, 2009028, 2010001, 2010016, 2102001, 2102020, 2103001, 2105028, 2106001, 2108011, 2108017, 2109022, 2110009, 2110030, 2111023, 2111045, 2112026, 2203020, 2204016, 2204035, 2207014, 2207046; 61105/February 28, 2025/00643169063815/2009060, 2011017, 2102011, 2102027, 2102047, 2103002, 2104042, 2105029, 2108001, 2108012, 2108016, 2109023, 2110010, 2111021, 2111024, 2111046, 2112027, 2202034, 2203034, 2204027, 2206008, 2207015, 2207033; 61106/December 31, 2024/00643169063822/2010017, 2103003, 2105030, 2105038, 2108013, 2108018, 2109004, 2110011, 2110031, 2111025, 2202035, 220302; 61110/March 31, 2025/00643169063839/2006039, 2010018, 2101014, 2102002, 2102028, 2103004, 2105031, 2108019, 2109024, 2110012, 2111026, 2201005, 2202036, 2204025, 2204040; 61111/April 30, 2025/00643169063846/2008039, 2010019, 2102021, 2103005, 2105032, 2106004, 2108014, 2110013, 2111027, 2202037, 2203022, 2203028, 2207016; 62100/January 31, 2025/00643169063853/2007021, 2008040, 2009026, 2009061, 2010014, 2011018, 2012001, 2101015, 2101020, 2102012, 2102022, 2102048, 2103019, 2104043, 2105033, 2106005, 2108020, 2109001, 2109020, 2109028, 2110014, 2110032, 2110033, 2111028, 2111035, 2111037, 2112001, 2112002, 2202038, 2203023, 2204001, 2204028, 2206013, 2206027, 2207017, 2207048, 2208009; 62105/January 31, 2025/00643169063860/2008005, 2008041, 2009029, 2010015, 2010020, 2012002, 2101016, 2102003, 2102013, 2102029, 2103006, 2103036, 2104044, 2105036, 2106035, 2108003, 2108015, 2109002, 2109003, 2110003, 2110015, 2110016, 2111022, 2111033, 2111047, 2201006, 2202014, 2202025, 2203015, 2203025, 2203031, 2204023, 2204036, 2204041, 2206014, 2206028, 2207018, 2207044; 62106/December 31, 2024/00643169063877/2008042, 2010021, 2101017, 2102023, 2103007, 2104045, 2105039, 2106007, 2108004, 2108021, 2109005, 2110017, 2110034, 2111029, 2202039, 2207019, 2207051; 62110/March 31, 2025/00643169063884/2006041, 2007022, 2008006, 2008043, 2010022, 2101009, 2101018, 2102026, 2102049, 2103020, 2104046, 2105001, 2105040, 2108005, 2108022, 2110004, 2110018, 2111036, 2201007, 2202015, 2202031, 2204017, 2204042, 2205031, 2205032, 2206007, 2207020, 2207049; 62111/May 31, 2025/00643169063891/2010023, 2012003, 2101019, 2102024, 2103008, 2105034, 2106009, 2108023, 2109006, 2110019, 2110035, 2111030, 2111034, 2202040, 2203024, 2204002, 2207021, 2207037, 2208027; 62158/April 30, 2025/00643169063907/2011019, 2103009, 2105035, 2110020, 2111031, 2202041, 2203033, 2204029, 2207006
Medtronic Neurosurgery is recalling Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE due to Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may present clinically with signs and symptoms of an acute inflammatory process, comparable to infection.
Recommended Action
Per FDA guidance
On 6/13/23 recall notices were mailed to customers and distributors and they were asked to do the following: 1) Return affected product to the recalling firm. 2) Complete and return the Customer Confirmation Form to neuro.quality@medtronic.com 3) This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Customers with questions can contact Customer Service at 800-874-5797 Option 1.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.