Medtronic Vascular Rashkind Balloon Septostomy Catheter, 6F, Recess Balloon, Closed end, 50 cm, Pediatric, Single Lumen Product number: 008764 GTIN: 00613994760289 The Rashkind balloon septostomy catheters are designed for creating an atrial septal defect or enlarging an existing atrial septal defect as a treatment option for patients with a Cyanotic Congenital heart defect. Cyanotic congenital heart defects are rare conditions that are associated with high morbidity and mortality. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Rashkind Balloon Septostomy Catheter, 6F, Recess Balloon, Closed end, 50 cm, Pediatric, Single Lumen Product number: 008764 GTIN: 00613994760289 The Rashkind balloon septostomy catheters are designed for creating an atrial septal defect or enlarging an existing atrial septal defect as a treatment option for patients with a Cyanotic Congenital heart defect. Cyanotic congenital heart defects are rare conditions that are associated with high morbidity and mortality.
Brand
Medtronic Vascular
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Lot Numbers : GFCP2154 GFCR2591 GFCR2592 GFCS2156 GFCU2199 GFCV2587 GFCW1652 GFCW2431 GFCX2712 GFCY2997 GFDN0246 GFDN3553 GFDN3554 GFDN3555 GFDP2922 GFDQ2500 GFDR2919 GFDS2209 GFDS2210 GFDT2188 GFDT2189 GFDT2204 GFDT2205 GFDU2170 GFDU2171 GFDV2179 GFDV2180 GFDV2181 GFDW2390 GFDW2391 GFDW2392 GFDX2711 GFDX2712 GFDY1135 GFDY1136 GFDZ1692 GFDZ1693 GFEP1181 GFEP1182 GFEP1183 GFEP1184 GFER2080 GFER2081
Medtronic Vascular is recalling Rashkind Balloon Septostomy Catheter, 6F, Recess Balloon, Closed end, 50 cm, Pediatric, Single Lumen due to Subsequent failure of catheters in the field and failed quality testing. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Subsequent failure of catheters in the field and failed quality testing
Recommended Action
Per FDA guidance
An immediate field corrective action was initiated by Medtronic field representatives. Customers were notified verbally to quarantine and return affected product on 25-Aug-2020. The following information was verbally provided to affected customers by Medtronic representative: "A supplier for Rashkind Catheter product has notified Medtronic that it has shipped product to Medtronic that may contain an issue. The supplier conducts a routine test in production. In two separate lots, the manufacturing specification was not met by all samples, but the lots were inadvertently shipped to Medtronic. We are requesting that you immediately refrain from using this product and place it in quarantine until further notice. You will be receiving an official recall notice with instructions for product return. Your Sales Representative will assist you with this process. We apologize for the inconvenience; however this action is necessary to ensure the utmost patient safety with the use of our products" In the United States on 09-Sep-2020, Medtronic began delivering an "Urgent Medical Device Recall" letter via UPS 2- day delivery to risk managers who, according to Medtronic records have received affected product. The notification states the following: 1. Identify and quarantine all unused Rashkind Balloon Catheters as listed in notification 2. Return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at 1-888-283-7868 to initiate a product return. Your local Medtronic Representative can assist you in the return of this product. 3. Please complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. Even if you do not have any product to return 4. Please forward the notice to all those who need to be aware within your organization.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.