Medtronic Xomed, Inc. VARI-STIM III Nerve Locator/Stimulator, Sterile, Medtronic Xomed, Inc. The Vari-Stim III nerve locator/stimulator is a sterile, single-use disposable, hand-held, battery-operated device intended to stimulate motor nerves exposed during surgery for the purpose of identification or location. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VARI-STIM III Nerve Locator/Stimulator, Sterile, Medtronic Xomed, Inc. The Vari-Stim III nerve locator/stimulator is a sterile, single-use disposable, hand-held, battery-operated device intended to stimulate motor nerves exposed during surgery for the purpose of identification or location.
Brand
Medtronic Xomed, Inc.
Lot Codes / Batch Numbers
Lot Number 58536200.
Products Sold
Lot Number 58536200.
Medtronic Xomed, Inc. is recalling VARI-STIM III Nerve Locator/Stimulator, Sterile, Medtronic Xomed, Inc. The Vari-Stim III nerve loc due to Device may fail to operate or operate intermittently during use and such operation could result in a lack of nerve stimulation when nerve tissue is ac. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Device may fail to operate or operate intermittently during use and such operation could result in a lack of nerve stimulation when nerve tissue is actually contacted by the stimulator.
Recommended Action
Per FDA guidance
All consignees were contacted by telephone (initiated on November 10, 2008). After telephone contact, each consignee was sent a follow-up fax/letter, "Urgent Product Recall Notification". The letter described the issue and actions for customers to take in response to the recall. Customers were instructed to check their inventories and return affected product to Medtronic Xomed, Inc. Customers were also instructed to fax their completed "Customer Contact and Product Return Checklist" to Medtronic Xomed, Inc. at 1-904-296-2386. Direct questions to Medtronic ENT Customer Service at 1-800-874-5797 (select Option 1 for Customer Service) and ask for the Vari-Stim Field Action Contact Person.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.