Neo-Tee T-Piece Resuscitator (Mercury Medical) – Flow Controller Defect (2024)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842
Brand
Mercury Enterprises, Inc. dba Mercury Medical
Lot Codes / Batch Numbers
Part/UDI-DI/Lot: 1050805/10641043508053, 30641043508057/2421450805, 2423550805, 2423650805, 2426750805, 1050808/10641043508084, 30641043508088/2426450808, 2426750808, 2426850808, 2429050808, 2429350808, 1050809/10641043508091, 30641043508095/2426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809, 1050810 /10641043508107, 30641043508101/2426750810, 2429350810, 2429750810, 2429850810, 2430250810, 1050811/10641043508114, 30641043508118/2426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811, 1050814/10641043508145, 30641043508149/2426450814, 2426750814, 2429050814, 2429750814, 2430250814, 1050832/10641043508329, 30641043508323/2429050832, 242985083, 1050839/10641043508398, 30641043508392/2430350839, 1050840/10641043508404, 30641043508408/2429050840, 2429750840, 1050841/10641043508411, 30641043508415/2429050841, 1050842/10641043508428, 30641043508422/2426750842
Products Sold
Part/UDI-DI/Lot: 1050805/10641043508053, 30641043508057/2421450805, 2423550805, 2423650805, 2426750805; 1050808/10641043508084, 30641043508088/2426450808, 2426750808, 2426850808, 2429050808, 2429350808; 1050809/10641043508091, 30641043508095/2426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809; 1050810 /10641043508107, 30641043508101/2426750810, 2429350810, 2429750810, 2429850810, 2430250810; 1050811/10641043508114, 30641043508118/2426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811; 1050814/10641043508145, 30641043508149/2426450814, 2426750814, 2429050814, 2429750814, 2430250814; 1050832/10641043508329, 30641043508323/2429050832, 242985083; 1050839/10641043508398, 30641043508392/2430350839; 1050840/10641043508404, 30641043508408/2429050840, 2429750840; 1050841/10641043508411, 30641043508415/2429050841; 1050842/10641043508428, 30641043508422/2426750842
Mercury Enterprises, Inc. dba Mercury Medical is recalling Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1 due to Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the hi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
Recommended Action
Per FDA guidance
On 12/10/2024, "URGENT: FIELD SAFETY NOTICE" letters were mailed and emailed to medical facilities and distributors who were asked to do the following: Medical Facilities/Distributors: 1) Users should cease use and distribution of affected product and immediately quarantine the affected product. 2) Contact firm's customer service department at uscustomerservice@mercurymed.com (Domestic) and InternationalCustomerService@mercurymed.com (International) to assist you on how to send the affected product back to the firm for product replacement or credit. 3) Please notify all personnel subject to using these devices. 4) Complete and return the response form via email to regulatoryaffairs@mercurymed.com In addition, distributors were asked to do the following: 5) Provide a copy of this recall notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you. If you have any questions regarding this action, contact the firm at 800-237-6418/727-573-0088, or e-mail at regulatoryaffairs@mercurymed.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026