Microaire Surgical Instruments Wire Driver Coupler, Model 7100-045, a component of MicroAire Drill/Reamer Model 7500. Driver couplers are reuseable medical devices, non-sterile, packed one per box and labeled in part ***MicroAire Surgical Instruments, 1641 Edlich Drive, Charlottesville, VA 22911***. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Wire Driver Coupler, Model 7100-045, a component of MicroAire Drill/Reamer Model 7500. Driver couplers are reuseable medical devices, non-sterile, packed one per box and labeled in part ***MicroAire Surgical Instruments, 1641 Edlich Drive, Charlottesville, VA 22911***.
Brand
Microaire Surgical Instruments
Lot Codes / Batch Numbers
Serial numbers: 1007, 1008, 1009, 1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019, 1020, 1021, 1022, 1024, 1025, 1026, 1027, 1028, 1029, 1030, 1031, 1032, 1033, 1034, 1035, 1036, 1037, 1038, 1039, 1042, 1043, 1044, 1045, 1047, 1049, 1050, 1051, 1052, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1064, 1065, 1066, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076, 1079, 1082, 1083, 1084, 1087, 1088, 1089, 1090, 1094, 1095, 1096, 1097, 1099, 1100, 1101, 1102, 1103, 1104, 1105, 1106, 1107, 1108, 1109, 1110, 1112, 1113, 1114, 1115, 1117, 1118, 1119, 1120, 1121, 1122, 1124, 1126, 1127, 1128, 1129, 1130, 1132, 1133, 1134, 1135, 1136, 1137, 1138, 1139, 1140, 1141, 1142, 1143, 1144, 1145, 1146, 1162, 1163, 1164, 1165, 1166, 1167, 1168, 1169, 1170, 1171, 1172, 1173, 1175, 1176, 1177, 1178, 1179.
Products Sold
Serial numbers: 1007, 1008, 1009, 1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019, 1020, 1021, 1022, 1024, 1025, 1026, 1027, 1028, 1029, 1030, 1031, 1032, 1033, 1034, 1035, 1036, 1037, 1038, 1039, 1042, 1043, 1044, 1045, 1047, 1049, 1050, 1051, 1052, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1064, 1065, 1066, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076, 1079, 1082, 1083, 1084, 1087, 1088, 1089, 1090, 1094, 1095, 1096, 1097, 1099, 1100, 1101, 1102, 1103, 1104, 1105, 1106, 1107, 1108, 1109, 1110, 1112, 1113, 1114, 1115, 1117, 1118, 1119, 1120, 1121, 1122, 1124, 1126, 1127, 1128, 1129, 1130, 1132, 1133, 1134, 1135, 1136, 1137, 1138, 1139, 1140, 1141, 1142, 1143, 1144, 1145, 1146, 1162, 1163, 1164, 1165, 1166, 1167, 1168, 1169, 1170, 1171, 1172, 1173, 1175, 1176, 1177, 1178, 1179.
Microaire Surgical Instruments is recalling Wire Driver Coupler, Model 7100-045, a component of MicroAire Drill/Reamer Model 7500. Driver couple due to Wiring pin locking mechanism defective, potential to come apart in surgical site.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Wiring pin locking mechanism defective, potential to come apart in surgical site.
Recommended Action
Per FDA guidance
The firm notified consignees by certified letter 9/23/03. The recall notifications were serial number specific and advised consignees of the potential defect with the wirning pin/locking mechanism. Consignees were instructed to cease use and distribution and obtain a return authorization for repair. Distributors were requested to extend the recall and recover and return devices for repair.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026