YST Comprehensive QC Set (Microbiologics) – labeling error (2020)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Microbiologics QC SETS AND PANELS, YST Comprehensive QC Set, REF 5195P, containing Trichosporon dermatis, ATCC 204094, QR Code: (01)70845357030770, (17)200831, (10)5195-04.
Brand
Microbiologics Inc
Lot Codes / Batch Numbers
Catalog Number: 5195P (KWIK-STIK kit that contains 0778P). Lot Numbers: 5195-04, 5195-05. UDI 70845357030770.
Products Sold
Catalog Number: 5195P (KWIK-STIK kit that contains 0778P). Lot Numbers: 5195-04, 5195-05. UDI 70845357030770.
Microbiologics Inc is recalling Microbiologics QC SETS AND PANELS, YST Comprehensive QC Set, REF 5195P, containing Trichosporon derm due to KWIKSTIK 2 Pack (REF 0778P) and QC Set and Panels: KWIKSTIK YST Comprehensive QC Set (REF 5195P) contain product labeled as Trichosporon dermatis (T. . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
KWIKSTIK 2 Pack (REF 0778P) and QC Set and Panels: KWIKSTIK YST Comprehensive QC Set (REF 5195P) contain product labeled as Trichosporon dermatis (T. dermatis) which may contain Candida Utilis instead of T. dermatis.
Recommended Action
Per FDA guidance
Microbiologics notified customer on 02/12/2020 via "Urgent Medical Device Correction" letter. The recall letter was emailed to consignees and the consignees were instructed to properly discard any affected products they have. The recall letter identified the affected product and lot numbers. Reason for the correction was also noted in the recall letter.