Microgenics Corporation QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3
Brand
Microgenics Corporation
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Model/Part Number: 10015573 UDI Codes: (01) 0884883 01259 2 (17) 220310 (10) 74175462 (01) 0884883 01259 2 (17) 220512 (10) 74267412 Lot Numbers: 74175462 74267412 Calibrators within the set: QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3
Microgenics Corporation is recalling QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIM due to Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result valu. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
Recommended Action
Per FDA guidance
On 01/20/2022, the firm contact via email its international customers regarding that the QMS Tacrolimus Calibrator was recovering between 82 to 94% of its target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay. On 03/10/2022, the firm notified its U.S. customer of the issue. Customers were instructed to: 1)Determine if they are using or have inventory of QMS Tacrolimus Calibrator catalog number 10015573, lot 74175462 or lot 74267412. 2)Discontinue use of affected in-date lots, and transition to new lot(s) where corrections have been made. Lot 74395275 and Lot 74459127 have had the testing correction applied and are currently available in stock to replace any of their remaining inventory of lots 74175462 or 74267412. 3)Dispose of the effected lots in accordance with local, regional, national regulations. 4)Contact your health care professional, as appropriate, for further action. " This notice needs to be passed on to all who need to be aware within your organization or to any organization where the potentially affected calibrator lots have been transferred. " Tacrolimus results reported with these affected calibrator lots do not need to be reanalyzed. " Retain a copy of this letter for your laboratory records. " Call Technical Service if assistance is needed for lot replacement (free of charge). " Please complete the Acknowledgment & Receipt Form attached to this Notification within 5 days of the date of this letter to Thermo fisher Scientific Technical Service. For questions, contact: -FAX: 888-527-8001 (USA only) or +1-510-979-5420 (outside USA) -PHONE: (+1)-800-232-3342 then press option 2, then option 2, or contact your local country representative. -E-mail: TechService.mgc@thermofisher.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AZ, CA, FL, ID, IL, IA, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC, TN, TX, VA, DC