Product
Synergetics I Pack Injection Kit- Tray for use with Ophthalmic injections Model/Catalog Number: 18060
Brand
MICROspecialties, Inc.
Lot Codes / Batch Numbers
UDI: 10841305101064 LOT: P62764979R
Products Sold
UDI: 10841305101064 LOT: P62764979R
MICROspecialties, Inc. is recalling Synergetics I Pack Injection Kit- Tray for use with Ophthalmic injections Model/Catalog Number: 18 due to Could not confirm products were exposed to the proper sterilization cycle can potentially cause post operative wound infection or injection site infec. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Could not confirm products were exposed to the proper sterilization cycle can potentially cause post operative wound infection or injection site infection.
Recommended Action
Per FDA guidance
MICROspecialties issued Voluntary recall for Synergetics Eye Trays letter on 6/3/24 to the Distributor Synergetics. Letter states reason for recall, health risk and action to take: Review: Inventory for affected product codes Communicate with customers impacted by this notice. Segregate and Quarantine all affected product. Utilize return instructions below Disseminate this notice to any customers who may have been affected. Return the below acknowledgement form by Fax to:860-788-2030 Or email to: quality@microspecialties.com Contact quality@.microspecialties.com for product return instructions. Please return by 6/25/24. For additional information regarding this please contact quality@.microspecialties.com.