Miltex, Inc. Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.
Brand
Miltex, Inc.
Lot Codes / Batch Numbers
Lot number S07C06
Products Sold
Lot number S07C06
Miltex, Inc. is recalling Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually package due to Blade may break: Over milling of the blade resulted in tip more narrow than specified.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Blade may break: Over milling of the blade resulted in tip more narrow than specified.
Recommended Action
Per FDA guidance
An Urgent Medical Device Recall letter dated 11/17/08 with an accompanying fax back form was sent to each customer. Each distributor was instructed to review their inventory, complete the form, and return any affected product. The distributors were also instructed to conduct a sub recall notification.