Mindray DS USA, Inc., dba Datascope Patient Monitoring SmarTemp Probe Covers for use with the VS-800 Monitor; Rx only, Mindray DS USA, Inc. Mahway, NJ 07430. Packaged in boxes of 20, Model ST010, also distributed in boxes of 2000 (10 individual boxes of 200); SmarTemp Module Kit (contains one box of 20 probes) and SmarTemp Temperature Probe Kits (contains one box of 20 probes). Probe Cover for use with the SmarTemp Reusable Temperature Probe. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SmarTemp Probe Covers for use with the VS-800 Monitor; Rx only, Mindray DS USA, Inc. Mahway, NJ 07430. Packaged in boxes of 20, Model ST010, also distributed in boxes of 2000 (10 individual boxes of 200); SmarTemp Module Kit (contains one box of 20 probes) and SmarTemp Temperature Probe Kits (contains one box of 20 probes). Probe Cover for use with the SmarTemp Reusable Temperature Probe.
Brand
Mindray DS USA, Inc., dba Datascope Patient Monitoring
Lot Codes / Batch Numbers
All lots of the following product numbers: 1. M09A-20-62124 Covers, Temp Probe- box of 20 probes and M09A-30-62128 Covers
Products Sold
All lots of the following product numbers: 1. M09A-20-62124 Covers, Temp Probe- box of 20 probes and M09A-30-62128 Covers, Temp Probe - box of 2000 (10 boxes of 200) 2. A box of 20 Disposable Probe Covers (part number M09A-20-62124) are also included in the following kits: 0020-00-0192 - SmarTemp Module Kit and 6006-30-39590 - SmarTemp Temperature Probe Kit.
Mindray DS USA, Inc., dba Datascope Patient Monitoring is recalling SmarTemp Probe Covers for use with the VS-800 Monitor; Rx only, Mindray DS USA, Inc. Mahway, NJ 0743 due to SmarTemp Disposable Probe covers may contain microscopic holes in the molded seam that would allow liquids or body fluids inside the cover and potenti. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
SmarTemp Disposable Probe covers may contain microscopic holes in the molded seam that would allow liquids or body fluids inside the cover and potentially contaminate the temperature probe.
Recommended Action
Per FDA guidance
Mindray DS USA, Inc. issued an "Urgent Product Recall" notification dated December 31, 2009 by certified mail. Consignees were informed of the affected product and requested to discard all remaining stock. Replacement product will be provided by the firm at no cost. For further information, contact Mindray DS USA, Inc. at 1-201-995-8000.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.