Hardware configuration of the BeneVision Central Monitori... (Mindray DS USA, Inc. dba Mindray North America) – when the worstation of the benevision... (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: Audio Driver version 6.0.9484.1 HP Pro Mini 400 G9 Desktop PC with Realtek HD Audio Driver Officially Released by HP (audio driver version 6.0.9484.1)
Brand
Mindray DS USA, Inc. dba Mindray North America
Lot Codes / Batch Numbers
Benevision Mini PC Workstation. Part Number E115-050935-00. Quantity 4
Products Sold
Audio Driver Version: 6.0.9484.1. UDI-DI: 06936415972417. HP Pro Mini 400 G9 Desktop PC. Realtek HD Audio Driver Released by HP. 1. WS host package (mini PC). Part Number 115-050935-00. Quantity 902 2. Exch, Benevision Mini PC Workstation. Part Number E115-050935-00. Quantity 4
Mindray DS USA, Inc. dba Mindray North America is recalling Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Dis due to When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.
Recommended Action
Per FDA guidance
Mindray notified consignees of the recall on about 11/18/2025 via certified mail. Consignees were informed that they may continue to use the Central Monitoring System. If the HP Pro Mini 400 G9 Desktop PC used by the WS experiences the audio failure or screen freezes, restart the HP Pro Mini 400 G9 Desktop PC. Consignees were instructed to ensure applicable clinical staff are aware of the issue and letter and shared with any customer if further distributed. Consignees were also asked to complete and return the Acknowledgement and Receipt form provided. Mindray will provide a patch update for the audio driver of the affected HP Pro Mini 400 G9 Desktop PCs located at each facility and is anticipated to take approximately 10-15 minutes per HP Pro Mini 400 G9 Desktop PC. A representative from Mindray's Care team will contact consignees to schedule the update.
Verify this information on the official source