Naviscan PET Systems PEMView Software used with Naviscan PEM Flex Solo I and PEM Flex Solo II PET Scanner, for nuclear medicine imaging. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
PEMView Software used with Naviscan PEM Flex Solo I and PEM Flex Solo II PET Scanner, for nuclear medicine imaging.
Brand
Naviscan PET Systems
Lot Codes / Batch Numbers
Serial numbers: C3, P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12, P14, P002-B-070119-01, P002-B-070316-01, P002-B-070423-01, P002-B-070531-01, P002-B-070629-01, P002-B-070712-01
Products Sold
Serial numbers: C3, P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12, P14, P002-B-070119-01, P002-B-070316-01, P002-B-070423-01, P002-B-070531-01, P002-B-070629-01, P002-B-070712-01
Naviscan PET Systems is recalling PEMView Software used with Naviscan PEM Flex Solo I and PEM Flex Solo II PET Scanner, for nuclear me due to Naviscan has identified a potential safety issue with all versions of PEM View Software commonly used for analyzing patient images acquired on the PEM. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Naviscan has identified a potential safety issue with all versions of PEM View Software commonly used for analyzing patient images acquired on the PEMFIex product line. 1. The Positron Emission Mammography Standardized Uptake Value (PEMSUV) and distance measurements for the product mentioned above may be inaccurate if a zoomed image has been opened, measurement tools are used, and the applet rema
Recommended Action
Per FDA guidance
Telephone notification regarding issues #1 and #2 were carried out on August 1 and 2, 2007 and included instructions for a validated work-around to avoid the software issues. Written notifications with these same validated work-around instructions were sent on August 2, 2007 via certified mail. Customers were notified that Naviscan has identified a potential safety issue with all versions of PEM View Software. Issue #3 was identified after the notification to customers for issues #1 and #2. Telephone notification regarding issues #3 was carried out via telephone and certified mailing beginning August 16, 2007. Customers were informed that Naviscan is preparing software to resolve these issues and that they would be contacted by Naviscan Customer Support to ensure they thoroughly understand the issues and the recommended procedure to prevent these issues from occurring. If customers have any questions they were instructed to contact Naviscan Customer Support at 1-888-628-4159 or service@navsicanpet.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.