Nellcor Puritan Bennett Nellcor Puritan Bennett brand Pulse Oximeter, Model NPB295, Made in Ireland; Distributed by Nellcor Puritan Bennett, Inc. Pleasanton, CA 94588 U.S.A. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Nellcor Puritan Bennett brand Pulse Oximeter, Model NPB295, Made in Ireland; Distributed by Nellcor Puritan Bennett, Inc. Pleasanton, CA 94588 U.S.A.
Brand
Nellcor Puritan Bennett
Lot Codes / Batch Numbers
All units with serial numbers beginning with G01844386.
Products Sold
All units with serial numbers beginning with G01844386.
Nellcor Puritan Bennett is recalling Nellcor Puritan Bennett brand Pulse Oximeter, Model NPB295, Made in Ireland; Distributed by Nellco due to The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.
Recommended Action
Per FDA guidance
Customers of record were notified of the advisory by letter sent via US Mail on October 18, 2004, via certified mail. A second letter was sent on December 9, 2004, advising customers that upgrades would be available. Upgrades would be available via returns to the firm''s service center, upgrade kits and speaker replacements via self-installation, and a third party service provider.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.