Nicolet Biomedical Div of Viasys Healthcare Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between December 2004 and March 2007 (Axiomtek Panel PC Model Number: 2147-671) VIASYS NeuroCare Part Numbers: 842-673800, 842-673900, 982A0421 and 982A0422. Manufactured, designed, developed and marketed by VIASYS NeuroCare Inc., Madison, WI 53711 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between December 2004 and March 2007 (Axiomtek Panel PC Model Number: 2147-671) VIASYS NeuroCare Part Numbers: 842-673800, 842-673900, 982A0421 and 982A0422. Manufactured, designed, developed and marketed by VIASYS NeuroCare Inc., Madison, WI 53711
Brand
Nicolet Biomedical Div of Viasys Healthcare
Lot Codes / Batch Numbers
ON040060, ON050120, ON050130, ON050168, ON050186, ON050197, ON050199, ON050203, ON050206, ON050212, ON050220, ON050231, ON060232, ON060243, ON060254, ON060256, ON060257, ON060269, ON060270, ON060272, ON060275, ON060276, ON060277, ON060278, ON060279, ON060280, ON060288, ON060290, ON060291, ON060292, ON060299, ON070301, ON070308.
Products Sold
ON040060, ON050120, ON050130, ON050168, ON050186, ON050197, ON050199, ON050203, ON050206, ON050212, ON050220, ON050231, ON060232, ON060243, ON060254, ON060256, ON060257, ON060269, ON060270, ON060272, ON060275, ON060276, ON060277, ON060278, ON060279, ON060280, ON060288, ON060290, ON060291, ON060292, ON060299, ON070301, ON070308.
Nicolet Biomedical Div of Viasys Healthcare is recalling Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel due to Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.
Recommended Action
Per FDA guidance
Certified direct mail delivery of URGENT MEDICAL DEVICE FIELD CORRECTION (RECALL) were made to consignees on 9/11/07 by Nicolet Biomedical Div of Viasys Healthcare. The letter describes the product and problem. VIASYS requests that customers cease use of the video option of the device until the monitor can be replaced by VIASYS authorized personnel. Letter also requests consignees to fax or mail back the response form.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026