Northwest Medical Physics Equipment Inc ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy treatment system. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy treatment system.
Brand
Northwest Medical Physics Equipment Inc
Lot Codes / Batch Numbers
ISOLOC Software Version 6.0, Part Number NT-NW-425-520
Products Sold
ISOLOC Software Version 6.0, Part Number NT-NW-425-520
Northwest Medical Physics Equipment Inc is recalling ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy trea due to When ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect.
Recommended Action
Per FDA guidance
Software to fix the described problem was written and customers contacted by e-mail on 12/30/2004 that a software patch, titled 12302004 was available to their customers on the firm's internet website.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026